|
Abstracts from the Sixth EPUAP
Open Meeting (continued)
Budapest 2002
Pressure ulcer prevention in x-ray departments
Audrey Brown
Derry, UK
Purpose
The time a patient spends on an X-ray table is increasing as procedures
become more complex with increasing interventional radiology, therefore
there is a growing need to look at the risk of pressure damage occurring.
It has been suggested but not proven that a patient in this hospital developed
a pressure sore during a prolonged Computed Tom-ography (CT) Guided Biopsy,
which was noticed on return to the ward. This project aimed to assess
the situation.
Method
This prospective study collected data from 80 patients, 20 patients on
each of the four different mattresses/support surfaces used in our department.
Data was collected to assess each patients risk of pressure damage and
a risk assessment score determined using the Braden Scale. A skin inspection
of eleven pressure areas pre and post examination was documented and duration
of examination noted.
Results
53.8% acquired Stage 1 or –1 skin damage in the immediate post examination
period. Logistic regression analyses indicated that the effectiveness
of the 4 mattresses differed. The closed foam mattress used for patients
undergoing angiography was most effective for preventing skin changes/worsening
of any kind. The risk of patient as measured by Braden Scale was found
to be significant. Relative to a patient judged at no risk, a high-risk
patient was 19.99 times as likely (P=0.0010) to exhibit worsening of any
kind and 40.36 times (P+0.004) as likely to exhibit worsening at the sacrum.
Conclusion
A potential risk appears to exist for pressure damage occurring in a radiology
department. The length of time a patient lies on the X-ray table mattress
is not the contributory factor but whether they are at risk of developing
pressure area damage as measured by a risk assessment tool. Difference
in type of mattress needs to be addressed.
Key words: Pressure sore, radiology, mattress/support surface, risk assessment,
evidence-based documentation, patient care.
Evaluation of alternating pressure air mattresses
using a time-based pressure threshold technique and laser Doppler micro-vascular
perfusion measurements on the heel
Shyam Rithalia
University of Salford, Salford, UK
Linda Russell
Queens Hospital, Derbyshire, UK
Introduction
Heels have substantially higher tissue interface pressures and are prone
to ulceration compared to other bony prominences. Although many different
type of APAMs are used for the prevention and treatment of pressure ulcers,
few high quality randomised controlled trials (RCTs) are available on
which to base purchasing decisions. Faced with this situation, physiological
measurements are increasingly being used as a surrogate. We have previously
reported a time-based technique1, which calculates pressure relief index
(PRI), for analysing the ability of such systems, and have demonstrated
that different designs produce variable results in this regard. The present
study compares the pressure relieving characteristics of two (Nimbus3,
Huntleigh Healthcare Ltd and Duo, Hill-Rom Ltd) using PRI and blood perfusion
measurements.
Methods
A total of eleven healthy adult volunteers (eight males, three females)
participated in the investigation for interface pressure and blood perfusion
measurements. They were recruited from postgraduate students and staff
of the University of Salford. Their written consent was obtained prior
to the commencement of the measurements in a temperature regulated (23
– 25°C) room. They were asked to lie supine on the bed, whilst
wearing normal light clothing. Care was used to place both left and right
heels on the centre of the same air cell of a mattress. Interface pressure
(IP) transducer (OPM, model II, Talley Group Ltd, Hants, UK) was placed
under the right heel and laser Doppler (LD) probe (Softflo, model BPM2,
Vasamedics Inc, MN, USA) on the left heel. Blood perfusion was reported
in arbitrary units (AU) following zero calibration. Measurements of IP
and LD were taken simultaneously over at least two alternating cycles,
as well as mattress air pressure. Differences between various pressures
and blood perfusion LD values over one cycle were analysed using Student’s
t-test or the Mann Whitney U-test depending on whether or not data were
normally distributed. A difference was considered significant when p <
0.05.
Results
The Nimbus3 provided consistently lower IPs during deflation phase of
the cycle (Nimbus3 = 20.2 ( 8.9 mm Hg and Duo = 68.5 (13.0mm Hg; p <
0.001). The time intervals calculated over 60 minutes when IP remained
below arbitrarily chosen thresholds of 30 and 20mm Hg were 10 and 3 minutes
respectively for the Nimbus. The Duo achieved zero time below either threshold.
There was no significant difference in peak pressures under the heel despite
the Duo offering significantly lower (P< 0.001) air cell pressures
in the heel region. Skin LD levels integrated over time were significantly
greater (p < 0.001) for the Nimbus (10171.3 ± 6721.5 AU) compared
with those for the Duo (6721.5 ± 1588.2 AU).
Summary
Based on the present combined results of IP and blood perfusion in healthy
subjects, it appears that, to produce hyperaemic reaction in the skin
tissue or re-perfusion after loading an appropriate course of action is
to provide periodic complete or near complete off-loading. A previously
published clinical study supports this finding.2 It is also important
to note that low air cell pressures do not necessarily produce lower IPs
under the heel, contrary to the intuitive classical notion.
Reference
Rithalia SVS, Gonsalkorale M. Journal of Rehabilitation Research and Development
1998; 35: 225– 230.
2. Russell L. Reynolds TM. Journal of Wound Care 2000; 9: 52–55.
Change of Body Pressure and Shear Force at Head
Up / Down Motion
Takehiko Ohura1, Makoto Takahashi2, Maki Mimura3 and Ryuji Kajiwara3
1) Pressure Ulcers and Wound Healing Research Center, Keishokai, Sapporo,
Japan
2) Biomedical Systems Engineering, Hokkaido University, Sapporo, Japan
3) Molten Corporation, Hiroshima, Japan
Introduction
Recently in Japan, pressure ulcers have been thought to be caused by the
factors: mechanical stress x time x frequency
rather than just the pressure of the body surface x time.
Until today it has been clinically difficult to measure direct shear force,
however, due to our design and development of a new electrical device
we have been able to measure shear force and surface pressure simultaneously,
allowing us to obtain some excellent data.
Method
1. Measurement of three points
1) Bone prominence of sacral vertebra
2) 25cm above from the middle of sacral vertebra
3) Middle of upper thigh.
2. Method
Subjects lay on a normal hospital mattress with a bed pad and a cotton
sheet. A sensor pad was directly applied to the subjects skin and then
dressed in cotton underwear. In both knees elevated and non-elevated positions
continuing measurements were taken of the upper body shear force when
raised from 30° through to 80°.
Results
1. As the upper body is raised the shear force on the sacral region changes
the direction between the upper and lower body.
2. Change of shear force was found in both knees elevated and non-elevated
positions before upper body was also raised.
3. Change of shear force was evident through the entire rotation between
the 30° and 80° upper- body elevated position.
4. Residual shear force was surprisingly detected when the upper body
was lowered.
Summary
The use of an electrical device developed in Japan, which measures the
shear force has allowed us to clinically obtain a great amount of critical
new data and evidence.
Numerical Accuracy of pressure mapping systems
– a comparative evaluation
P. Diesing, D. Hochman and U. Boenick
Technical University Berlin, Department of Mechanical Engineering, Berlin,
Germany
Introduction
Pressure mapping systems (PMS) has been established over the last decade
for to evaluate the influence of pressure on the development of pressure
ulcers (PU). Mat integrated sensor-arrays based on resistive and capacitive
effect became the state of the art against the older pneumatic systems.
Although the pressure measurement between the skin and the surface of
the bed or wheelchair is only an indicator for the pressure inside the
tissue, PMS seem to be the best way to evaluate the risk created by pressure
for PU in an region of interest. For to choose i.e. an optimal seat-cushion
for a wheelchair it might be enough, to compare the pictures of an PMS
to find the best cushion. For to define a lab-procedure to check pressure
relieving devices concerning their effectiveness to prevent PU it is necessary
to know the measurement errors creates by linearity, hysteresis, hammocking-effect
and effects of non sufficient resolution of the PMS. The main issue of
this paper is, to show the results of a short comparative evaluation of
the actual commercial offered PMS.
Methods
Four PMS (FSA, novel, Seatscan, Xsensor) had been evaluated at the TU
Berlin, Dep. Biomechanical Engineering. The systems had been calibrated
by the manufacturer directly before the mat was brought in. The systems
had been tested in regard to accuracy and linearity of a small area influenced
by the geometry of the force applicator, the hammocking-effect on different
surfaces (hard ground and two different foams) and hysteresis. Because
of the different resolutions of the sensormats we decided to compare calculated
normal forces instead of the measured pressure.
Results
Three systems (FSA, novel and Xsensor) has been integrated in the comparison.
The Seatscan-system was demonstrated with a 6 year old sensormat and the
linearity showed a curve which was not feasible, so the Seatscan system
was excluded from the comparison.
All systems showed a good linearity, even if there was a relevant difference
between mats with high resolution (32 x 32 sensors) and systems with low
resolution (16 x 16 sensors). The size of the applicator had an important
effect on the readings of the normal force. All PMS underestimated the
force applied on a small contact area. The minimum contact area for an
accurate measurement was in the range between 4.5cm2 and 19.6cm2. The
influence of the ham-mocking effect on the readings could be observed
as expected. The readings differed between the hard surface and a soft
foam underlay. The measured hysteresis curves differed in part from curves
measured in previous studies.
Summary
The evaluated PMS are excellent tools for the visualization of the pressure
in the daily use of a hospital. They allow in a very quick and very easy
way to find the area, where high pressures increase the risk of getting
an PU and how this pressure is influenced by different devices. The software
produces some analysis of the measured data in a comfortable way. However,
for the numerical measurement of high pressure peaks under bones PMS underestimate
the pressure, because of insufficient resolution and other effects. Additionally
the difference between the readings of different types of PMS will not
allow to compare values measured with different systems.
The implications of skin moisture retention during
the use of four contemporary wound dressings and two ointments
Yoshiteru Ishii, Kaoru Nishide* and Kiyomi Tomita*
Dept of Plastic surgery, Kenwakai Otemachi Hospital, Japan
*Smith & Nephew Japan
The moist wound environment is a well known concept in
the treatment of wounds. Many dressings reflect this concept and are used
to enhance healing. It is not clear to what degree the various dressings
are capable of influencing the moisture retention of the skin and whether
they vary in effectiveness. In addition, problems such as maceration of
the surrounding skin have been reported when using dressings that promote
a moist wound environment.
The aim of this study was to measure the ability of various dressings
to maintain a moist wound environment by investigating moisture retention
within the skin.
The moisture retention in the horny layer of the surrounding skin of 38
pressure ulcers was measured. Measurements were carried out within 1cm
of the wound area. The control was non-treated exposed skin away from
the wound area. Four types of wound dressing and two kinds of ointment
with gauze dressing were used on 38 wounds. The measurements were carried
out using Moisture Checker MY707S (Scalar Corp., Japan).
The dressings used were as follows: hydrocellular polyurethane foam (Allevyn™,
Smith & Nephew); hydropoly-mer (Tielle™, Johnson & Johnson);
hydrogel (Intrasite Gel™, Smith & Nephew); hydrofiber (Aquacel™,
Conva-Tec); bucladesine sodium ointment (Actosin™, Daiichi Pharm.);
and silver sulfadiazine cream (Geben Cream™, Mitsubishi-Tokyo).
Once the dressings were removed from the wound, the area was wiped with
a dry gauze and measured in the same spot three times. An average for
each spot was then calculated. The locations of the pressure ulcers were
as follows: 20 sacrum, 9 heel, 7 greater tochanter and 2 others.
The measurements of the surrounding skin area were compared with the average
moisture rate of the horny layers. The moisture retention score for each
dressing from the highest to the lowest was as follows: hydrofiber, hydrogel,
and polyurethane foam. For the hydropolymer dressing was the moisture
rate score was actually lower under the dressing than in the control area
(see table below).
| Product |
Test area
(under dressing, 1cm from wound) |
Control area
(non-treated exposed skin) |
Difference |
| Allevyn™ (n=26) |
43.3 |
31.0 |
12.3 |
| IntrasiteGel™ (n=4) |
47.2 |
33.7 |
13.4 |
| Aquacel™ (n=3) |
55.5 |
33.6 |
21.9 |
| Tielle™ (n=10) |
33.2 |
34.6 |
-1.6 |
| Actosin™ (Ointment) (n=3) |
54.5 |
34.6 |
19.9 |
| SSD™ (Geben Ointment) (n=32) |
52.6 |
32.5 |
20.1 |
| TOTAL n=38 |
44.2 |
32.8 |
11.4 |
Conclusion
The hydrocellular polyurethane foam dressing and the hydrogel appeared
to enable good moisture retention. The hydropolymer dressing appeared
to create a dry environment. These results were also evident from visual
inspection.
Strategies to identify the most appropriate dressing for a wound should
take into account the effect of the dressing on skin moisture retention
and the potential to dry out a wound or keep it too moist, and possibly
cause maceration.
|